• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC CR TIB INSERT STD, SZ 3, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC CR TIB INSERT STD, SZ 3, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-47-3013
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/10/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant devices (3823693) 02-010-03-0330 - logic cr femoral cem, right, sz 3; (4013214) 02-012-45-3040 - lgc tibial fit tray cem sz 3f/4t; (4237534) 200-02-35 - three peg patella 35mm.The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported via legal documentation that approximately 99 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear.No further issues or complications were reported.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR TIB INSERT STD, SZ 3, 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville 32653
3523771140
MDR Report Key20334079
Report Number1038671-2024-03742
Device Sequence Number8136095
Product Code JWH
UDI-Device Identifier10885862174369
UDI-Public10885862174369
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/09/2020
Device Catalogue Number02-012-47-3013
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2024
Initial Report FDA Received Date09/30/2024
Date Device Manufactured03/10/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-