| Catalog Number |
21-7052-24 |
| Medical Device Problem Codes |
Complete Blockage (1094); Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient was without drug for twenty minutes as the device was not working.Per reporter the device exhibited a "high priority alarm".The device had been changed to a new cassette that weekend, but it began alarming while the patient was in town.The home screen indicated a high priority alarm.Downstream occlusion and did not clear for twenty minutes.The reporter further noted that troubleshooting was performed by the patient where they changed to a different device, infusion line batch, cassette batch and the device in an effort to isolate the cause.The patient was able to get home and re-establish the infusion using a new system with no further issues.There was patient involvement, and no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone: (b)(6).H3: neither samples nor pictures were received for the analysis of this complaint the device history record of reported lot number: 4342624 was reviewed and it was confirmed that no non-conformities were reported during production.The quality inspection forms were reviewed, and it was observed that no defects were found, and the lot was released.Without the return of the samples a comprehensive failure investigation cannot be performed, and a probable cause cannot be determined.
|
| |
|