SMITH & NEPHEW, INC. 9.0MM FIXED ENDCUTTING RMR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Catalog Number |
71118226 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during the osteosynthesis procedure, the 9.0mm fixed endcutting rmr broke off when the medullary canal was finished.The procedure was resumed, without any delay, using a s+n back-up device.Oncologic patient, diagnosed with breast cancer, during the procedure went into cardiac arrest and after seven attempts of resuscitation died in the operating room due to natural causes, not due to a device related failure.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during the osteosynthesis procedure, the 9.0mm fixed endcutting rmr broke off when the medullary canal was finished.The pieces fell inside the patient.The device broke in two and did not remain inside the canal.It managed to get out by retracting the engine, and the piece was grabbed with rochester forceps.The procedure was resumed, without any delay, using a s+n back-up device.Oncologic patient, diagnosed with breast cancer, during the procedure went into cardiac arrest and after seven attempts of resuscitation died in the operating room due to natural causes, not due to a device related failure.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: additional information ¿h6: health effect - impact code.H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The photographs were reviewed, and did not provide any insight relative to the failure mode or the root cause of the alleged failure.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|