| Model Number |
GIF-N180 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed during the device inspection, that the gastrointestinal videoscope exhibited the nozzle of the insufflation channel clogged by an unidentified plastic material.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material is plastic material.However, we could not identify the foreign material.Additionally, no physical damage was confirmed at the place where the foreign material remained.There were no deviations from the instruction manual in the reprocessing steps at the material residue point.Therefore, a definitive root cause could not be determined.The events can be detected and prevented in accordance with the following ifu.Detection method is described in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection.Preventive measures are described in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|