Model Number GIF-N180
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was observed during the device inspection, that the gastrointestinal videoscope exhibited the nozzle of the insufflation channel clogged by an unidentified plastic material.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material is plastic material.However, we could not identify the foreign material.Additionally, no physical damage was confirmed at the place where the foreign material remained.There were no deviations from the instruction manual in the reprocessing steps at the material residue point.Therefore, a definitive root cause could not be determined.The events can be detected and prevented in accordance with the following ifu.Detection method is described in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection.Preventive measures are described in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20339551
Report Number9610595-2024-19582
Device Sequence Number8024275
Product Code FDS
UDI-Device Identifier04953170307508
UDI-Public04953170307508
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGIF-N180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2024
Supplement Date Received by Manufacturer11/14/2024
Initial Report FDA Received Date09/30/2024
Supplement Report FDA Received Date11/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-