ALCON RESEARCH, LLC - HUNTINGTON MONARCH II IOL DELIVERY SYSTEM, CARTRIDGE B; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977758
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A healthcare professional reported with a description of on implantation using the cartridge there was a small crack in the optic.The lens was left in the patient eye.Additional information was requested.
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.A qualified lens model/diopter and handpiece were indicated with a non-quailed viscoelastic.The cartridge was not returned.The reported lens damage could not be verified.The lens was left implanted.The root cause for the reported damage may be related to a failure to follow the ifu.A non-qualified viscoelastic was indicated.The ifu instructs: the iol delivery system is for implantation of qualified company foldable iols.The company cartridges are qualified for use with compatible company handpieces for the surgical implantation of company qualified foldable iols.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.It is unknown if adequate viscoelastic was used.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH II IOL DELIVERY SYSTEM, CARTRIDGE B
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20342972
Report Number1119421-2024-01936
Device Sequence Number15459757
Product Code MSS
UDI-Device Identifier00380659777585
UDI-Public(01)00380659777585(17)260522(11)230524(10)15655840
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K001157
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/22/2026
Device Catalogue Number8065977758
Device Lot Number15655840
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/02/2024
Supplement Date Received by Manufacturer11/22/2024
Initial Report FDA Received Date10/01/2024
Supplement Report FDA Received Date12/20/2024
Date Device Manufactured05/24/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACRYSOF MULTIPIECE IOL.; HEALON PRO 10MG/ML SODIUM HYALURONATE.; MONARCH III IOL DELIVERY SYSTEM, INJECTOR.
Patient Age83 YR
Patient SexMale
-
-