| Catalog Number |
MEGEN1 |
| Medical Device Problem Code |
Self-Activation or Keying (1557)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the bipolar megadyne working automatically without pressing a footswitch.Even when footswitch was not pressed it was working automatically.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 10/1/2024.D4: udi: the manufacturing date is currently not available.Therefore, the full udi is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Serial number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 12/12/2024 investigation summary per service manual operational and diagnostic analysis confirmed the self activation issue.The backplane pcb and footswitch pcb were identified in the investigation to address the issue.The repair is pending customer approval.The device history records were reviewed and certed by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/3/2025 corrected data: h11 investigation summary per service manual operational and diagnostic analysis confirmed the self activation issue.The backplane pcb was identified in the investigation to address the issue.The repair is pending customer approval.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/4/2025.Corrected data: report submission due date.
|
| |
|