DEPUY SPINE INC T HANDLE INSERTER; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Catalog Number |
PET30101T |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follo.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2024 during post-op inspection, several instruments were found to be worn or damaged.T-handle has a cross threaded inner driver attached.Trial won¿t load properly onto the trial handles.No complaints were noted during surgery.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: h4.H3, h4, h6: photo investigation: the product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photo investigation revealed that the t handle inserter no damage or cosmetic defects, and no observations pertaining to the nature of the reported event could be identified.Functionality of the device cannot be assessed through photo evidence provided.¿ as the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the t handle inserter would have contributed to the complained device issue.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Physical device investigation: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found it to be worn from the distal threads.The reported allegation can be confirmed.This issue can related to the unable to assemble allegations.The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test could not be performed as the mating device was not returned.The overall complaint was confirmed as the observed condition of the t handle inserter would contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): a review of the receiving inspection (ri) for t handle inserter was conducted identifying that lot number nn185356 was released in one batch.Batch 1: lot qty of (b)(4).Units was released on jan 10, 2023, with no discrepancies.Supplier: (b)(4).The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|