• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. NEXGEN COMPLETE KNEE SOLUTION; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. NEXGEN COMPLETE KNEE SOLUTION; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED Back to Search Results
Catalog Number 00597206532
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Synovitis (2094); Osteolysis (2377); Swelling/ Edema (4577)
Date of Event 09/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: item# 00596403210; lot# 65038731 item# 00598603702; lot# j6817421.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient underwent a revision approximately two (2) years and six (6) months post-implantation where only the patella and articular surface components were replaced.The physician suspected backside wear of the articular surface and tibia to be the cause.The patella also showed radiographic signs of bone lysis, so the surgeon wanted to replace it.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: a1; b4; b5; d2; g1; g3; g6; h1; h2; h10 d10: item# 00596403210; lot# 65038731 item# 00598603702; lot# j6817421 item# 00576401552; lot# 65022207 if any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual inspection of the returned patella exhibits signs of use (discoloration and wear) and bone cement remains.Primary notes suggest no intra operative complication.Revision notes states that patient was revised due to pain, swelling, synovitis.During the surgery, wear was seen underside of articular surface with some micro marking on the tibial component too.However, the tibial component was well fixed and was not explanted.Review of the device history record identified no deviations or anomalies during manufacturing.Root cause cannot be determined.This complaint is confirmed via product returned, operative notes and photographs provided.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN COMPLETE KNEE SOLUTION
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20344261
Report Number0002648920-2024-00304
Device Sequence Number12762429
Product Code JWH
UDI-Device Identifier00889024215146
UDI-Public(01)00889024215146(17)300117(10)65270022
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number00597206532
Device Lot Number65270022
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2024
Supplement Date Received by Manufacturer10/21/2024
03/04/2025
Initial Report FDA Received Date10/01/2024
Supplement Report FDA Received Date11/05/2024
03/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H11
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-