| Model Number |
KNOTLESS TENSIONABLE FIBERTAK BICEPS KIT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Insufficient Information (4580)
|
| Date of Event |
06/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 9/19/2024, an fda medwatch notification was received via email.A facility representative reported that the tip of the fibertak from an ar-3688 knotless fibertak biceps implant system broke off in the patient's humerus.This was discovered during a procedure in (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|