• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ9 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ9 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 157012165
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377)
Date of Event 03/19/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Medical records received ad 20 sept 2024.After review of medical records patient was revised due to failed internal prosthesis of the left hip.Operative notes indicated pain, metallosis, and found large area of osteolysis on posterior aspect of the femur.Doi: (b)(6) 2009 dor: (b)(6) 2024 left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).E3 initial reporter occupation: lawyer.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code:157012165, lot:c13ej1000 and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device product code:157012165, lot:c13ej1000 and no non-conformances / manufacturing irregularities were identified.H11 additional narrative: added: d10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ9 HI OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20345686
Report Number1818910-2024-20549
Device Sequence Number7880134
Product Code LPH
UDI-Device Identifier10603295060208
UDI-Public(01)10603295060208(17)180719(10)C13EJ1000
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K193398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/19/2018
Device Catalogue Number157012165
Device Lot NumberC13EJ1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/20/2024
Supplement Date Received by Manufacturer10/16/2024
Initial Report FDA Received Date10/01/2024
Supplement Report FDA Received Date10/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/21/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12/14 ARTICUL 40MM M SPEC+8.5; PINNACLE MTL INS NEUT40IDX58OD; PINNACLE SECTOR II CUP 58MM; SUMMIT DUOFIX TAP SZ9 HI OFF
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-