| Model Number |
VICMO 12.1 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).Manufacturer narrative - secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced low vault and refractive surprise.In a separate visit the lens was exchanged for a replacement lens of longer length and different diopter.The problem was resolved.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: lens was returned in micro-centrifuge vial, in liquid and residue/debris on product.Visual inspection found haptic broken.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|