• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JAGTOME RX 39; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JAGTOME RX 39; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00573080
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagtome rx 39 was used in the ampulla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the procedure, the cutting wire broke in half.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: the returned jagtome rx 39 was analyzed, and a visual evaluation noted that the cutting wire was broken and kinked from the proximal pierced hole, which are consistent with the findings when the device was observed under magnification.Additionally, the broken cutting wire were blackened.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, kinked and blackened.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagtome rx 39 was used in the ampulla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the procedure, the cutting wire broke in half.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAGTOME RX 39
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key20346524
Report Number3005099803-2024-04852
Device Sequence Number20846412
Product Code KNS
UDI-Device Identifier08714729776642
UDI-Public08714729776642
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00573080
Device Catalogue Number7308
Device Lot Number0034261488
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2024
Supplement Date Received by Manufacturer10/22/2024
Initial Report FDA Received Date10/01/2024
Supplement Report FDA Received Date11/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-