Catalog Number 71645060
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference number: case: (b)(4).H10: this complaint was opened by smith+nephew to document a patient complication identified through the review of clinical evidence from post market clinical data collection activities that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported, after an intramedulary nail revision performed on (b)(6) 2020 due to the backing out of the trigen low profile screw 5.0mm x 55mm and the breakage of the trigen low profile screw 5.0mm x 67.5mm.Later, the patient required an additional revision surgery on (b)(6) 2021 to replace the distal interlocking screws due to the breakage of the trigen low profile screw 5.0mm x 60mm.A new smith & nephew 57.5mm and 67.5 mm interlocking screws were implanted in exchange.
 
Additional Manufacturer Narrative
H10: h10: the information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case has already been reported to fda under the internal reference number (b)(4) and the mdr report number 1020279-2024-01657.The correct investigation and findings will be documented under (b)(4), mdr report number 1020279-2024-01657.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIGEN LOW PROFILE SCREW 5.0MM X 60MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20347929
Report Number1020279-2024-02005
Device Sequence Number7895740
Product Code HSB
UDI-Device Identifier00885556216194
UDI-Public885556216194
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K111025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71645060
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2024
Supplement Date Received by Manufacturer10/14/2024
Initial Report FDA Received Date10/01/2024
Supplement Report FDA Received Date10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71645075/TRIGEN LOW PROFILE SCREW 5.0MM X 75MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-