ARTHROCARE CORPORATION WEREWOLF FLOW 50 COBLATION WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 72290037
Medical Device Problem Codes Device Remains Activated (1525); Self-Activation or Keying (1557)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during a hip arthroscopy the flow 50 wand kept on going while no button was touched or the foot pedal was not used.So the probe has his on will.The procedure was completed with a smith and nephew back up device.There was a delay less than 30 minutes and no further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H11 h6: medical device problem code updated.
 
Additional Manufacturer Narrative
Section h6 was updated.Section h10: the reported device was received for evaluation.Visual inspection observed the returned instrument shows no manufacturing abnormalities.No visual issues.Product was out of the original packaging.No packaging returned.Functional evaluation revealed plugging in the wand causes a wand end of life error.Using a bypass box, the wand had no issues producing plasma or activating coagulation mode.The ablate button intermittently works when pressed.The handle was opened and found the button boards to be covered in liquid residue.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (the wand´s connector or cable got damaged, or saline/humidity entered the interior of the handle shorting the connection of the buttons).Based on this investigation, there is no evidence to suggest a design, material, or manufacturing issue.Therefore, the need for corrective action is not indicated.Internal complaint reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEREWOLF FLOW 50 COBLATION WAND
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20349489
Report Number3006524618-2024-00424
Device Sequence Number7880258
Product Code GEI
UDI-Device Identifier00885556617472
UDI-Public885556617472
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K162074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72290037
Device Lot Number2152899
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2024
Supplement Date Received by Manufacturer11/01/2024
01/29/2025
Initial Report FDA Received Date10/01/2024
Supplement Report FDA Received Date11/01/2024
02/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-