Model Number 12110
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that at an unspecified time, this dlp left heart vent catheters had an issue and does not retain its shape despite the presence of the metal rod.Failure to maintain the desired shape leads to a risk of heart injury (pulmonary vein, left atrium or left ventricle) in 2-year-olds.The problem recurred 2 times in a row and 8 cannulas were opened out of the 10 contained in a box, and all had the same defect.The surgeon had to change the batch number (not communicated) in order to complete the procedure successfully.There were no clinical consequences for the child., and no adverse patient effect reported with this event.
 
Additional Manufacturer Narrative
Update to b5: medtronic received information that prior to use of these dlp left heart vent catheters, it was reported that the devices did not retain their shape despite the presence of the metal rods.It was stated that a failure to maintain the desired shape could lead to a risk of heart injury (pulmonary vein, left atrium or left ventricle) in 2-year-olds.The problem recurred 2 times in a row and 8 cannulae were opened out of the 10 contained in a box, and all had the same defect.The surgeon had to change the batch number in order to complete the procedure successfully.There was no patient involvement, so no adverse effect occurred.Correction d4: expiration date added.Correction h4: device mfg date added.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DLP LEFT HEART VENT CATHETER
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key20350392
Report Number2184009-2024-00586
Device Sequence Number20860089
Product Code DWF
UDI-Device Identifier00643169880672
UDI-Public00643169880672
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K834039
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number12110
Device Catalogue Number12110
Device Lot Number2024030359
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2024
Supplement Date Received by Manufacturer10/03/2024
Initial Report FDA Received Date10/02/2024
Supplement Report FDA Received Date10/25/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-