| Model Number |
N/A |
| Medical Device Problem Codes |
Overheating of Device (1437); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2 ¿ foreign ¿ spain.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery the reciprocating saw overheated and stopped suddenly during use.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Review of the most recent checklist determined this device failed the rotation check due to the locking mechanism not moving when the drive is turned.The device was scrapped.Review of the device history record identified no deviations or anomalies during manufacturing.Review of complaint history identified additional similar complaints for the reported item and no additional similar complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|