MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH*PNEUMOPERITONEUM NDL; APPARATUS, PNEUMOPERITONEUM, AUTOMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH*PNEUMOPERITONEUM NDL; APPARATUS, PNEUMOPERITONEUM, AUTOMATIC Back to Search Results
Catalog Number PN120
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown robotic procedure, the safety needle remained exposed after puncture and did not retract.Opened a second product to complete case.No patient harm.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 10/2/2024.D4: batch # unk.Investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the pn120 device was returned with no apparent damage.During the visual inspection, the tip of the needle was noted with dried body fluids.The device was functionally tested to detect any issues.Upon evaluation of the device, it was functionally tested and passed.The device was fully functional according to the manufacturing requirements.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: during insertion, inspect the instrument handle to ensure that the ¿red safety indicator¿ slides proximally in the direction of the stopcock.This action retracts the blunt stylet and exposes the sharp needle tip for penetration.Do not attempt to use the endopath pneumoneedle if the blunt stylet does not retract into the needle sleeve.Once the blunt stylet is free of tension from the tissue, the ¿red safety indicator¿ resets itself back into the distal portion of the handle.Use appropriate testing to ensure that the endopath pneumoneedle is safely in the abdominal cavity.Once testing is completed, insufflate the abdomen.The event described could not be confirmed as no damages were observed and the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device lot 434a28 number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
APPARATUS, PNEUMOPERITONEUM, AUTOMATIC
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key20351135
Report Number3005075853-2024-07426
Device Sequence Number18155910
Product Code FDP
UDI-Device Identifier10705036012962
UDI-Public(01)10705036012962(17)260831(10)434A28
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K910875
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Catalogue NumberPN120
Device Lot Number434A28
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/05/2024
Initial Report FDA Received Date10/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-