• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP - SUPERDIMENSION INC EDGE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP - SUPERDIMENSION INC EDGE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number SDK4900-FT
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a procedure, after removing the working channel from the patient, the nurse noticed what appeared to be a very thin 'wire' at the tip of the catheter when it was flushed.The user tried to retrieve it, but it could not be grabbed and was somehow lost outside of the patient's body.The very thin wire appeared to come from inside the catheter.To resolve the issue, a new catheter was used.It was noted that no fragments were in the patient¿s body.The physician was able to complete the electromagnetic navigational bronchoscopy (enb) portion of the case.There was no patient injury.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: of kinked that has no relationship to the reported condition.The product analysis noted evidence that the device was not used as intended.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGE
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer (Section G)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20351147
Report Number3004962788-2024-00110
Device Sequence Number8053804
Product Code JAK
UDI-Device Identifier10884521545571
UDI-Public10884521545571
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K151376
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSDK4900-FT
Device Catalogue NumberSDK4900-FT
Device Lot Number529053
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2024
Supplement Date Received by Manufacturer01/10/2025
Initial Report FDA Received Date10/02/2024
Supplement Report FDA Received Date02/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
-
-