• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA25-90/I20-30
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Codes Aneurysm (1708); Clinical Endoleak (4843)
Date of Event 09/16/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft, and two (2) afx vela infrarenal's.Approximately five (5) years post initial procedure, during a routine follow up, there appeared to be a type iiib endoleak and the aneurysm had grown from 5.5cm to 7.5cm.The physician elected to reline the implanted afx2 bifurcated stent graft with a new afx2 bifurcated stent graft and the endoleak appeared to be resolved.During the deployment, after turning the yellow cap and pulling to release the wrapper, the wrapper would not come off.The issue was troubleshooted by pulling the contra yellow cover with significant force and pulling on the yellow cap strings with significant force at the same time.This released the wrapper.The patient is reported as doing well and expected to be discharged next day.
 
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the type iiib endoleak of the distal main body, aneurysm enlargement of 30 mm and additional endovascular procedure complaints are confirmed.This is consistent with the reported adverse event/incident.No contributing factors were identified.Device, user, procedure or anatomy relatedness of complaint could not be determined.No procedure related harms were identified.The final patient status was reported as discharged home on postoperative day one in good condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: date received by manufacturer has been updated.H6: result code: remove code 3233.H6: conclusion code: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key20351621
Report Number3011063223-2024-00141
Device Sequence Number18193165
Product Code MIH
UDI-Device Identifier00818009014825
UDI-Public(01)00818009014825(17)180526
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other,Recall
Type of Report Initial,Followup
Report Date (Section B) 09/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/26/2020
Device Model NumberBEA25-90/I20-30
Device Lot Number1697785011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/16/2024
Supplement Date Received by Manufacturer11/22/2024
Initial Report FDA Received Date10/02/2024
Supplement Report FDA Received Date12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/13/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA INFRARENAL, LOT #2012864002.; AFX VELA INFRARENAL, LOT #2260445002.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-