(h3) pending evaluation (d10) concomitant device(s): 320-15-01 - eq rev glenoid plate (b)(6).320-20-00 - eq reverse torque defining screw kit (b)(6).320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm (b)(6).320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm (b)(6).320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm (b)(6).320-20-38 - eq rev compress screw lck cap kit, 4.5 x 38mm (b)(6).321-20-00 - equinoxe reverse shoulder drill kit (b)(6).320-15-05 - eq rev locking screw (b)(6).
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c, d, e, g.The revision reported was likely due to disassembly between the humeral liner and humeral tray leading to subsequent unintended metal-on-meal articulation between components.However, migration of the liner subsequent to the disassembly, and the sequence of events could not be confirmed.Potential contributions of user and patient-related considerations to the event, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the disassembly of the humeral liner cannot be confirmed from the reported information, as the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|