| Catalog Number |
G136 |
| Medical Device Problem Codes |
Overheating of Device (1437); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While using a cavitron plus g136 they allege that they are having no water flow and overheating.No injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6) 2024 evaluation tech: (b)(4).W2h reg, needle valve, hp flow cntrl damaged continuous water leaking due to the solenoid not closing completely.Missing battery door, damaged water flow control on the handpiece cable.Top cover and cabinet has broken body posts.No batteries.Added handpiece as a precaution for any damages sustained from overheating will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.No handpiece received for evaluation.
|
| |
|