Model Number IBI-81104
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 09/10/2024
Type of Reportable Event Serious Injury
Event or Problem Description
During an atrial fibrillation procedure, a pericardial effusion occurred in the right atrium requiring a pericardiocentesis.Following transseptal puncture, prior to inserting the volt catheter the pericardial effusion was observed with the ice catheter.Protamine was administered, and when performing the pericardiocentesis, the blood was noted to be venous in nature.The patient is stable.It is unknown what caused the pericardial effusion; however, it could have been caused by maneuvering the catheter in the coronary sinus or by maneuvering the introducer for transeptal puncture.There were no alleged performance issues with any abbott devices.
 
Additional Manufacturer Narrative
Additional information: g3, h2, h3, h6, h11.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL
2375 morse ave
irvine CA 92614
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key20354558
Report Number2030404-2024-00065
Device Sequence Number3492650
Product Code DRF
UDI-Device Identifier05414734302975
UDI-Public(01)05414734302975(10)10360614(17)270531
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberIBI-81104
Device Catalogue NumberIBI-81104
Device Lot Number10360614
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2024
Supplement Date Received by Manufacturer11/22/2024
Initial Report FDA Received Date10/02/2024
Supplement Report FDA Received Date12/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight85 KG
-
-