ST. JUDE MEDICAL INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Model Number |
IBI-81104 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
09/10/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, a pericardial effusion occurred in the right atrium requiring a pericardiocentesis.Following transseptal puncture, prior to inserting the volt catheter the pericardial effusion was observed with the ice catheter.Protamine was administered, and when performing the pericardiocentesis, the blood was noted to be venous in nature.The patient is stable.It is unknown what caused the pericardial effusion; however, it could have been caused by maneuvering the catheter in the coronary sinus or by maneuvering the introducer for transeptal puncture.There were no alleged performance issues with any abbott devices.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h2, h3, h6, h11.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
|
| |
|