| Model Number |
9010 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Fever (1858); Unspecified Infection (1930); Pain (1994); Pseudoaneurysm (2605)
|
| Date of Event |
04/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 12/2025) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical trial that approximately three days post index procedure using a drug-coated balloon catheter in the basilic vein, the subject had an adverse event of fever, pain and swelling at anastomosis.It was further reported that a large pseudoaneurysm was noted on ultrasound.Reportedly, the adverse event was possibly related to the device but definitely related to the procedure and av access circuit.Reportedly, the adverse event doesn't require av access circuit re-intervention.The current status of the patient is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical trial that approximately three days post an index procedure using a drug-coated balloon catheter in the basilic vein, the subject allegedly had an adverse event of fever, pain, and swelling at anastomosis.It was further reported that a large pseudoaneurysm was noted on ultrasound.Reportedly, the adverse event was possibly related to the device but definitely related to the procedure and arteriovenous access circuit.Evidently, the adverse event does not require arteriovenous access circuit re-intervention.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, d4 (medical device model #), g3, h6 (patient, component, method) section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical trial that approximately three days post an index procedure using a drug-coated balloon catheter in the basilic vein, the subject had an adverse event of infected pseudoaneurysm.It was further reported that the adverse event was possibly related to study device and definitely related to procedure and av access circuit.Reportedly, the adverse event doesn't require av access circuit re-intervention.The current status of the patient was unknown.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical trial that approximately three days post index procedure using a drug-coated balloon catheter in the basilic vein, the subject had an adverse event of pseudoaneurysm at the anastomosis site.It was further reported that the adverse event was possibly related to study device and definitely related to procedure and av access circuit.Reportedly, the adverse event doesn't require av access circuit re-intervention.It was further reported that the subject had an adverse event of bacteremia.The event was not related to the device but definitely related to the procedure and av access circuit.The event did not result in an av access circuit re-intervention.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, g3, h6 (patient).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|