| Catalog Number |
TRS100SB2 |
| Medical Device Problem Code |
Material Perforation (2205)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The customer reported that the trs100sb2, anchor tissue retrieval system was being used on (b)(6) 2024 during a lap chole procedure when it was reported ¿2 of the anchor tissue bags (trs100sb2) that were defective.¿.Further assessment questioning found that ¿one bag would not close, the other bag tore when trying to close it.It was also confirmed that the bag did not fragment or fall into the surgical site.A 3rd bag was used to remove the specimen¿.There was no report of injury, medical intervention, or extended hospitalization to the patient or user.This report is being raised on the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned, and no photographic evidence was provided.Therefore, the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review shows a total of 3 devices for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 21 complaints, regarding 23 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised that only physicians trained in the techniques of laparoscopic or minimally invasive surgery and the use of retrieval bags to remove tissue should use the product.Do not use the anchor¿ tissue retrieval system¿ if resistance is met upon deployment.Improper use of the anchor¿ tissue retrieval system¿ may result in perforation of tissue and subsequent bleeding.Care should be taken when using sharp and electro-surgical devices.Note the size of the trocar cannula when removing a specimen through the trocar cannula and seal system.If required, enlarge the port site to aid removal of the anchor¿ tissue retrieval system¿ by conmed.We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The customer reported that the trs100sb2, anchor tissue retrieval system was being used on (b)(6) 2024 during a lap chole procedure when it was reported ¿2 of the anchor tissue bags (trs100sb2) that were defective.¿.Further assessment questioning found that ¿one bag would not close, the other bag tore when trying to close it.It was also confirmed that the bag did not fragment or fall into the surgical site.A 3rd bag was used to remove the specimen¿.There was no report of injury, medical intervention, or extended hospitalization to the patient or user.This report is being raised on the reported malfunction with potential for injury upon reoccurrence.
|
| |
|