| Model Number |
CR2 |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer will receive a replacement device.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not emit audio prompts.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient, if needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The product assessment center (pac) technician evaluated the device was able to verify the reported issue.An analysis of the electronic records showed that the issue was caused by a depleted battery.The battery became depleted due to the continuous beeping from the device being in not ready-call service status.It was observed that the device went to not ready status due to service code.But a root cause of the service code could not be determined.The customer received a replacement device.The customer's device was archived by stryker.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not emit audio prompts.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient, if needed.There was no patient use associated with the reported event.
|
| |
|
Search Alerts/Recalls
|