BIOMET 3I T3¿ PRO TAPERED IMPLANT 4/3MM (D) X 8.5MM (L); DENTAL IMPLANT Back to Search Results
Catalog Number T3PT4385
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Pain (1994)
Date of Event 08/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Multiple reports are being submitted for this event.A4: patient weight is not provided / unknown.E1: initial reporter¿s title is not provided / unknown.
 
Event or Problem Description
It was reported that sites 12 and 13 has an allergic reaction 7 months post op.Noted pain site grafted with allograft.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D4: unique device identifier (udi) number and expiration date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H4: device manufacture date was updated.H6: type of investigation codes were added: 4109, 4111, 4114 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie did not receive the reported implant for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and one (1) other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are customer error and patient factors (material selection.) therefore, based on the available information, device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3¿ PRO TAPERED IMPLANT 4/3MM (D) X 8.5MM (L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20358556
Report Number0001038806-2024-02250
Device Sequence Number8126710
Product Code DZE
UDI-Device Identifier00844868046752
UDI-Public(01)00844868046752(17)270823(10)2022071481
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K213672
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberT3PT4385
Device Lot Number2022071481
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/03/2024
Supplement Report FDA Received Date02/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-