| Model Number |
P870D |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report from a baxter technician to report that resident bed frame ac power cable was broke with exposed copper wires.The bed was located at the account.There was no patient/user injury, medical intervention, symptom, or adverse event reported.This report was filed in our complaint handling system as complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the ac power cord needed to be replaced.Per the hillrom service manual, it is necessary for the resident ltc bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Make sure the power cord is not damaged and it is connected to the bed.Replace the power cord as necessary.Check the ac inlet to psm cable assembly for kinks, damage, and connections.Replace or connect the cable as necessary.Replace parts as necessary.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the ac power cord to resolve the reported event.Based on this information, no further action is required.
|
| |
|