• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA MEDIAL CONGRUENT ARTICULAR SURFACE LEFT 12MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA MEDIAL CONGRUENT ARTICULAR SURFACE LEFT 12MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Loss of Range of Motion (2032); Scar Tissue (2060); Joint Contracture (4528)
Date of Event 09/05/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a manipulation under anesthesia to address arthrofibrosis accompanied by stiffness, limited range of motion and a 1-5 degree flexion contracture approximately three (3) months following left knee arthroplasty.The patient was reportedly improving following the manipulation.Reportedly, no intraoperative complications were noted during the initial procedure.
 
Additional Manufacturer Narrative
(b)(4).D10 - concomitant devices - persona cruciate retaining narrow femoral component left size 6 catalog#: 42508006001 lot#: 66266110, persona cemented keel tibia left size d catalog#: 42536006701 lot#: 65635671, persona cemented all poly patella 29mm catalog#: 42540000029, lot#: 66600814, palacos rg 1x40 single bone cement catalog#: 00111314001, lot#: 68911257.The complainant has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b7, g3, g6, h2, h6, h11.The device history records were not reviewed as the reported event was determined to be unrelated to zimmer biomet devices.Arthrofibrosis is defined as the development of fibrous scar tissue within or surrounding a joint.Arthrofibrosis is a known post-operative procedure-related complication that can occur from surgical implantation of a new joint replacement as well as from previous injuries or surgical procedures.Scar tissue formation is a normal healing response; however, the buildup of such can result in pain, stiffness, limited range of motion, and difficulty properly ambulating.If excess scar tissue develops, conservative measures such as exercises or physical therapy would be attempted first.If these attempts fail, surgical intervention such as manipulation under anesthesia, arthroscopic arthrolysis, or open arthrotomy would become necessary to remove the fibrous tissue and restore joint function.The root cause of the reported event was determined to be related to the procedure rather than zimmer biomet devices.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA MEDIAL CONGRUENT ARTICULAR SURFACE LEFT 12MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20363413
Report Number3007963827-2024-00386
Device Sequence Number14401755
Product Code MBH
UDI-Device Identifier00889024468993
UDI-Public(01)00889024468993(17)290122(10)66337787
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42512100412
Device Lot Number66337787
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2024
Supplement Date Received by Manufacturer10/16/2024
Initial Report FDA Received Date10/03/2024
Supplement Report FDA Received Date10/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Hospitalization; Required Intervention; Required Intervention;
Patient SexFemale
Patient Weight86 KG
Patient EthnicityHispanic
-
-