This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added: h3, h4, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is possible that a high-energy endo therapy accessory (laser, high frequency, etc.) was used without ensuring a sufficient distance from distal end.However, we could not specify the cause of the suggested event.A definitive root cause could not be determined.Detection method is described in ifu: gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection_3.3 inspection of the endoscope as follows: 8 inspect the entire distal end of the endoscope including the objective lens and light guide lens for any irregularities such as scratches, chips, cracks, stains, discoloration, deformation, and gaps around the lens.There are cautions when high-energy endotherapy accessories are used in ifu: gif/cf/pcf-190 series operation manual_4.3 using endotherapy accessories.Never emit high-frequency current before confirming that the distal end of the high-frequency endotherapy accessory is in the endoscope¿s field of view.Also, confirm that the electrode section and the mucous membrane in the vicinity of the target area are at an appropriate distance from the distal end of the endoscope.If the high-frequency current is emitted while the distal end of the endotherapy accessory is not visible or too close to the distal end of the endoscope, patient injury, bleeding, and/or perforation as well as equipment damage can result.Olympus will continue to monitor field performance for this device.
|