| Catalog Number |
305759 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
08/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd needle eclipse 25x5/8 safety shield broke off.It was reported by customer that they are we are having some safety issues with our bd needles.On multiple occasions the pink safety shield has popped off and also when we are screwing it on to the vials sometimes it breaks off.This happened to me personally before and yesterday it happened to one of our other lvns who actually had a needle stick incident.This is really dangerous working with our pediatric patients because sometimes they have fast hands too and i would hate for one of them to be poked.Verbatim: rcc received a complaint via email.Email(s) attached.Item: 305759.Quantity affected: (b)(4).Serial/lot number:(b)(6).Po : (b)(4).Item: 305759.Quantity affected: (b)(4).Serial/lot number: (b)(6).Po :(b)(4),reported issue: we are having some safety issues with our bd needles.On multiple occasions the pink safety shield has popped off and also when we are screwing it on to the vials sometimes it breaks off.This happened to me personally before and yesterday it happened to one of our other lvns who actually had a needle stick incident.This is really dangerous working with our pediatric patients because sometimes they have fast hands too and i would hate for one of them to be poked.
|
| |
|
Additional Manufacturer Narrative
|
|
Our quality engineer inspected the 200 representative samples submitted for evaluation.The reported issue of safety mechanism failure was not confirmed upon inspection of the samples.Analysis of the samples showed no damages or defects.Samples passed the safety shield inspection and the activation/ locking force test.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this lot meets our manufacturing specification requirements.The root cause could not be determined as the defect was not replicated.
|
| |
|
Event or Problem Description
|
|
Material #:305759; batch#:4023388.It was reported by customer that they are we are having some safety issues with our bd needles.On multiple occasions the pink safety shield has popped off and also when we are screwing it on to the vials sometimes it breaks off.This happened to me personally before and yesterday it happened to one of our other lvns who actually had a needle stick incident.This is really dangerous working with our pediatric patients because sometimes they have fast hands too and i would hate for one of them to be poked.Verbatim: rcc received a complaint via email.Email(s) attached.Item: 305759; quantity affected: (b)(4); serial/lot number: (b)(6); po : (b)(6).Item: 305759; quantity affected: (b)(4); serial/lot number: (b)(6); po : (b)(6).Reported issue: we are having some safety issues with our bd needles.On multiple occasions the pink safety shield has popped off and also when we are screwing it on to the vials sometimes it breaks off.This happened to me personally before and yesterday it happened to one of our other lvns who actually had a needle stick incident.This is really dangerous working with our pediatric patients because sometimes they have fast hands too and i would hate for one of them to be poked.
|
| |
|