MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MAR +4 10D 32IDX48OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MAR +4 10D 32IDX48OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121932148
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the acetabular insert ruptured.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : rupture of the acetabular insert.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed that the rim of the pinn mar +4 10d 32idx48od has three out of six anti-rotation device (ard) tabs sheared off.The fractured fragments were not returned for evaluation.Additionally, the liner appeared to have been edge loaded causing eccentric deformation in the rim of the device and ultimately contributing to the failure of the anti-rotation devices (ards).Based on the observations of the pinn mar +4 10d 32idx48od, it is suspected to conclude that a disassociation event occurred between the liner and the cup.There is no indication that a design or manufacturing issue has caused or contributed to the reported event.Although the liner failure cannot be traced to design or manufacturing, a definite root cause cannot be established due to there being multiple factors that may influence.Consideration must be given to all potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.It is highly recommended to the patient to consult with their healthcare professional for further assessment.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.A manufacturing record evaluation was performed for the finished device product description: pinn mar +4 10d 28idx46od product code: 121928146 lot number: jy2754 and no non-conformances or manufacturing irregularities were identified.The overall complaint was confirmed as the observed condition of the pinn mar +4 10d 32idx48od would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device product description: pinn mar +4 10d 32idx48od product code: 121932148 lot number: jh1126 and no non-conformances or manufacturing irregularities were identified.Device history review : a manufacturing record evaluation was performed for the finished device product description: pinn mar +4 10d 32idx48od product code: 121932148 lot number: jh1126 and no non-conformances or manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 32IDX48OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key20367806
Report Number1818910-2024-20817
Device Sequence Number3795906
Product Code LPH
UDI-Device Identifier10603295014034
UDI-Public(01)10603295014034(17)260630(10)JH1126
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K033273
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Catalogue Number121932148
Device Lot NumberJH1126
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/23/2024
Supplement Date Received by Manufacturer10/31/2024
11/28/2024
Initial Report FDA Received Date10/04/2024
Supplement Report FDA Received Date11/01/2024
11/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexUnknown
-
-