|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The microsensor (id (b)(6)) was returned for evaluation.Device history record (dhr) - the lot 7121744 conformed to the specifications when released to stock.Failure analysis - the catheter was inspected.It was noted that the stopcock connected to the catheter y-fitting was installed improperly.The stopcock could not be tightened (screwed) onto the y-fitting as configured.A syringe with water was connected to the catheter, the catheter tip was clamped, and pressure was applied with the syringe.No water was observed leaking from the catheter.Unable to confirm the complaint.Root cause analysis - the complaint was not confirmed in the complaint investigation and failure analysis.The potential root causes include surgeon applies sutures/ligatures too tight perforating catheter or misuse of device.
|