|
Updated fields: d4, d8, d9, g6, h1, h2, h3, h4, h6, h8, h11.The microsensor catheter kit (826653) was returned for evaluation.Failure analysis - the catheter was inspected.No defects were observed.A syringe with water was connected to the catheter, the catheter tip was clamped, and pressure was applied with the syringe.No water was observed leaking from the catheter.Root cause - the complaint was not confirmed in the complaint investigation and failure analysis evaluation.Potential root causes include surgeon applies sutures/ligatures too tight perforating catheter or mishandling of device.
|