• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE MECHANICAL LITHOTRIPTOR V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE MECHANICAL LITHOTRIPTOR V Back to Search Results
Model Number BML-V442QR-30
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.Full e1 address establishment name: (b)(6) hospital.
 
Event or Problem Description
It was reported the single use mechanical lithotriptor distal guidewire tip broke.The issue occurred during a therapeutic endoscopic retrograde cholangiopancreatography procedure.The procedure was completed using a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon of "guidewire tip torn" was presumed to have been due to the guide wire distal end did not meet strength specifications due to suboptimal molding conditions during manufacturing.The instruction manual contains a warning to the user regarding this event: (drawing number: gk5911, revision number: 23).·when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire while holding the distal tip as shown in figure 4.20.Be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as shown in figure 4.21.This may damage the distal tip.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE MECHANICAL LITHOTRIPTOR V
Common Device Name
SINGLE USE MECHANICAL LITHOTRIPTOR
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20369579
Report Number9614641-2024-01923
Device Sequence Number3745763
Product Code LQC
UDI-Device Identifier04953170218422
UDI-Public04953170218422
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
CLASS2EXMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBML-V442QR-30
Device Lot Number41K
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2024
Supplement Date Received by Manufacturer11/15/2024
Initial Report FDA Received Date10/04/2024
Supplement Report FDA Received Date11/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
G-2602545A VISIGLIDE2 SINGE USE GUIDEWIRE
Patient Age80 YR
Patient SexMale
-
-