| Model Number |
320574 |
| Medical Device Problem Code |
Misconnection (1399)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Consumer reported finding the non patient end hard to insert into the pen.Consumer reported finding the patient end not sharp and painful to insert into site.Lot # 3179397; catalog# 320574; date of event unknown; sample status awaiting sample unused.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was not able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
| |
|