The lifevent evo2 was returned to the third-party service center and device evaluation completed on 24 july 2025 with the following findings: the alleged device error code e-059 was found on the device, but was not able to be reproduced during evaluation and testing.This ventilator service required code indicates the internal or detachable li-ion battery is not charging due to a hardware fault for greater than, or equal to, one minute.The unit arrived for evaluation with no detachable battery.While testing the unit with a detachable battery attached, the error code e-59 could not be reproduced resulting in the conclusion that the customer's detachable battery was defective and should be replaced.This device has been serviced, inspected, tested, met acceptance criteria in accordance with all of the applicable customer requirements.Additional findings not related to the malfunction/issue: one in-line filter was noted to be contaminated and required replacement.The air inlet foam filter is replaced due to preventative maintenance.
|