|
The lifevent evo2 was returned to the third-party service center and device evaluation completed on 24 july 2025 with the following findings: the alleged device error code e-059 was found on the device but was not able to be reproduced during evaluation and testing.This ventilator service required code indicates the internal or detachable li-ion battery is not charging due to a hardware fault for greater than, or equal to, one minute.The unit arrived for evaluation with no detachable battery.While testing the unit with a detachable battery attached, the error code e-59 could not be reproduced.This device has been serviced, inspected, tested, met acceptance criteria in accordance with all of the applicable customer requirements.Additional findings not related to the malfunction/issue: one in-line filter was noted to be contaminated and required replacement.The air inlet foam filter is replaced due to preventative maintenance.
|