| Model Number |
MR28010721 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Catheter rupture (lumen).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A follow up report will be submitted when the investigation is complete.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
The device involved in the incident was not returned for evaluation.A drawing of the catheter was provided indicating the fracture occurred at the lumen to hub extension junction.Without an evaluation of the actual device involved a definitive root cause could not be determined.The device was implanted 14 days prior to the issue.The contract manufacturer conducted a review of the manufacturing records for the lot number provided.The device history record review established the device was manufactured within specification and customer requirements.There were no deviations or process changes in the manufacturing and/or materials for this part number.All inspections and sequence of operations were properly followed and documented.A 100% leak test is performed during inspection of this device.This test is capable of detecting lumen holes/tears in the lumen that could lead to a malfunction.
|
| |
|
Search Alerts/Recalls
|