BECTON, DICKINSON AND COMPANY (BD) BD PRESET¿ ECLIPSE¿; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 364391
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: "material #: 364391; lot/batch #: 4095709.Bd received 1 sample and 2 photos for investigation.The photos were reviewed and the customer¿s indicated failure mode for leakage was observed.Additionally, the customer sample was evaluated by functional draw testing and the indicated failure mode for leakage with the incident lot was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode leakage.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
 
Event or Problem Description
It was reported while using a bd preset¿ eclipse¿ syringe, blood leaked past the plunger stopper.A new blood sample was collected from the patient.This occurred with three (3) devices.There was no report of adverse impact to patients or users.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRESET¿ ECLIPSE¿
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key20371443
Report Number9617032-2024-01605
Device Sequence Number17857496
Product Code JKA
UDI-Device Identifier50382903643917
UDI-Public(01)50382903643917(11)240404(17)260331(10)4095709
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K202446
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2026
Device Catalogue Number364391
Device Lot Number4095709
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2024
Initial Report FDA Received Date10/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-