| Model Number |
WA4KL100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the telescope had a broken lens.The issue occurred during an unspecified procedure.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The customer's allegation was confirmed.Three attempts were performed to obtain additional information, but no response was received from the customer.Based on the results of the investigation, it is likely that the following led to the malfunction: a large mechanical force due to a shock, fall or collision with other equipment.A definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|