MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507445MC |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
05/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: right breast prosthesis rupture, area of palpable concern on left breast.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision procedure with mentor memorygel breast implant 445cc gel breast prostheses when the right breast prosthesis ruptured post implantation.Rupture of the patient¿s right breast prosthesis was diagnosed via mammogram, ultrasound, and mri.Additionally, the mri indicated an area of palpable concern in the left breast.It was reported to be a probably benign mass.As a result, the patient underwent bilateral explantation and replacement with mentor memorygel breast implant 475cc gel breast prostheses on (b)(6) 2024.This medwatch report is for the patient¿s left breast prosthesis.Refer to manufacturer report number 1645337-2024-10664 for the report for the patient¿s right breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On 09-oct-2024, the mentor failure analysis lab received the device for evaluation.On 11-oct-2024, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information reported, the patient underwent an explantation of the breast implant.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant was found to have a tear on the posterior view measuring approximately 9.8 cm.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the tear found, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|