|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Corrections added to fields: d5, d9, e2, g2.Additional information added to fields: d8, h4, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.The device was not returned to olympus for inspection; therefore, the customer's complaint was not confirmed.Based on the results of the investigation, the reported error description is most likely due to the effect of an excessive mechanical force; however, a definitive root cause could not be established as the device was not returned to olympus.Olympus will continue to monitor field performance for this device.
|