• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER Back to Search Results
Model Number MR850AEA
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A distributor in china reported that the audible alarm of an mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
 
Additional Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).Method: the subject mr850 respiratory humidifier was received at our fisher & paykel (f&p) regional office in china where it was inspected by a trained f&p service technician.Our investigation is based on the information provided by the f&p service technician, information provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: performance testing of the subject mr850 respiratory humidifier confirmed that no sound was generated from the speaker.Conclusion: the root cause of the speaker fault was not determined.Our mr850 respiratory humidifier product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is equipped with visual alarm indicators in addition to the audible alarm.
 
Event or Problem Description
A distributor in china reported that the audible alarm of an mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
RESPIRATORY HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key20380234
Report Number9611451-2024-00712
Device Sequence Number9592540
Product Code BTT
UDI-Device Identifier09420012407177
UDI-Public(01)09420012407177(10)2102524544(11)230221
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K073706
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMR850AEA
Device Catalogue NumberMR850AEA
Device Lot Number2102524544
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2024
Supplement Date Received by Manufacturer10/13/2024
Initial Report FDA Received Date10/06/2024
Supplement Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/21/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-