| Model Number |
53401 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The external pulse generator (epg) which was returned for preventive maintenance subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: at analysis it was determined that the external pulse generator (epg) battery board assembly was out of specification electrical failing functional testing.All found defective parts were replaced and all other identified issues were resolved.The instrument then passed all final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis: further investigation was performed on the external pulse generator(epg) printed circuit board (pcb), it was determined following testing that the pcb passed all automated tests.The pcb was found to be working correctly.Correction fdr code h6 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|