| Catalog Number |
254500543 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the trials had one missing spring.It was pointed out by the sterile processing department after the case.It was unknown if they came off in surgery or while being washed.These were used multiple times throughout the week but it was just noticed that these were missing.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that the trials had one missing spring.It was pointed out by spd after the case.It was unknown if the came off in surgery or while being washed.These were used multiple times throughout the week but were just made aware that these were missing.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the spring of the attune rp ps artic surf sz3 missing.The device shows signs of normal use of service.The failure mode is consistent with inserting an extractor device in between the trial and mating shim, and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface as well as the mating shim.The attune intuition surgical technique 0612-10-512 (page 53), emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, on page 51 of the surgical technique and per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.In the event of instrument breakage during use, ensure that all device fragments that may have entered the surgical site are removed prior to completion of the procedure, as patient injury may result.A dimensional inspection was not performed due to post manufacturing damage.A functional evaluation was not performed due to post manufacturing damage.The overall complaint was confirmed as the observed condition of the attune rp ps artic surf sz3 would contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
|
| |
|