MAUDE Adverse Event Report: SYNTHES GMBH 2.0 RAPID RESRB OBL L-PL 6H X 4H/RT-S; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.0 RAPID RESRB OBL L-PL 6H X 4H/RT-S; PLATE, BONE Back to Search Results
Catalog Number 852.264.01S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Physical Asymmetry (4573)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This complaint is from a literature source.The following literature cite has been reviewed: tomomatsu n, takahara n, akaike y, sato y, kurasawa y, yoda t.Postoperative stability of bioresorbable plates made of 85:15 poly (l-lactide-co-glycolide) in le fort i osteotomy.Oral surg oral med oral pathol oral radiol.2023 nov;136(5):569-576.Doi: 10.1016/j.Oooo.2023.04.016.Epub 2023 may 1.Pmid: 37640562.The study aimed to compare the postoperative stability of rapidsorb plates, rapidsorb combined with titanium (moj plates), and moj plates in patients who underwent le fort i osteotomy at tokyo medical and dental university hospital.The use of rapidsorb in the maxilla is a load-sharing application and therefore constitutes an approved indication.Discrepancies in the maxillary positions were measured using postoperative computed tomography data at 1 week and 1 year using the centroid method 3-dimensionally.Treatment with rapidsorb alone showed a more vertical discrepancy in the maxilla treatment with moj and rapidsorb+moj.The rapidsorb4 group was subdivided into 2 groups (under and over 1.0-mm) based on the change in the maxillary centroid.Between april 2018 and march 2021 were selected and evaluated retrospectively for cases using rapidsorb and titanium screws and plates.Lot, model and catalog number are not available, but the suspected depuy synthes device possibly associated with reported adverse events: rapidsorb plates (depuy synthes) made of 85.15 poly (l-lactide-co-glycolide) adverse event(s) and provided interventions possibly associated with unk - resorbable implants (qty 21) (n=20) patients in rap4 group had a poor stability, no intervention provided.(n=1) patient had a postoperative deviation by fixation with four rapidsorb plates is due to the tooth being somewhat vulnerable to occlusal loads, due to a decrease in the gripping strength of the screw caused by slight wobbling during tapping, no intervention provided.Adverse event(s) and provided interventions possibly associated with lot numbers were plate 6l87402, 6l71797, 6l58763, 6l39795 (qty 1) (n=1).A 24-year-old male had a plate fracture, in which a fracture fragment of rapidsorb was ejected from the wound 6 months after surgery.The fractured fragments have already been processed and are no longer present.There was no maxillary bone movement, and the patient had a good follow-up.Adverse event(s) and provided interventions possibly associated with lot number screw 4l95040 (qty 1) (n=1) a 24-year-old male had a plate fracture, in which a fracture fragment of rapidsorb was ejected from the wound 6 months after surgery.The fractured fragments have already been processed and are no longer present.There was no maxillary bone movement, and the patient had a good follow-up.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h4, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device history review: part:852.264.01s lot:6l87402 manufacturing site: werk bio oberdorf.Supplier: na.Release to warehouse date: 27 may 2020.Expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information received states that none of the patients experienced adverse events such as maxillary sinusitis or obvious nonunion.There was one case of plate fracture.The lot numbers were plate 6l87402, 6l71797, 6l58763, 6l39795 and screw 4l95040.The fractured fragments have already been processed and are no longer present.The slight postoperative deviation by fixation with four rapidsorb plates is due to the tooth being somewhat vulnerable to occlusal loads, and we believe this is due to a decrease in the gripping strength of the screw caused by slight wobbling during tapping.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0 RAPID RESRB OBL L-PL 6H X 4H/RT-S
Common Device Name
PLATE, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key20384210
Report Number8030965-2024-12547
Device Sequence Number7944270
Product Code JEY
UDI-Device Identifier10887587064307
UDI-Public(01)10887587064307(17)230401(10)6L87402
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Catalogue Number852.264.01S
Device Lot Number6L87402
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/19/2024
Supplement Date Received by Manufacturer10/15/2024
Initial Report FDA Received Date10/07/2024
Supplement Report FDA Received Date10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexMale
Patient Weight65 KG
-
-