• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR INC. SAFESHEATH® II INTRODUCER SET; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC. SAFESHEATH® II INTRODUCER SET; INTRODUCER, CATHETER Back to Search Results
Model Number SS7
Medical Device Problem Code Material Too Rigid or Stiff (1544)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the hubs of the safesheaths did not tear appropriately when they were being removed once the leads were in position.There were no adverse outcomes for this event.The hubs were eventually able to be split and torn, and the leads stayed in position during safesheath removal.No other information available.
 
Additional Manufacturer Narrative
Conclusion not yet available, evaluation in process.A follow-up report will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Additional Manufacturer Narrative
The following sections were updated in follow-up 1.The device was used for treatment.Rma issued for the device return.However, on 15jan2025 customer stated that device is not returning.No product was provided for analysis.Procedure completed successfully.There was no adverse event with the patient reported.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.No product was returned to oscor inc.For evaluation; however, a complaint notification was provided.Without the return of the actual device in question for evaluation, oscor inc.Is unable to determine the exact cause for this incident.At this time, it is not possible to assign a definitive root cause for the event as reported.The complaint is non-verifiable as the product was not returned for evaluation.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESHEATH® II INTRODUCER SET
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
OSCOR INC.
4875 palm harbor blvd.
palm harbor 34683
Manufacturer (Section G)
OSCOR INC.
4875 palm harbor blvd.
palm harbor 34683
Manufacturer Contact
ronnie greco
4875 palm harbor blvd.
palm harbor, FL 34683
7279372511
MDR Report Key20384433
Report Number1035166-2024-00108
Device Sequence Number17805353
Product Code DYB
UDI-Device Identifier30891492002228
UDI-Public30891492002228
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K122084
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSS7
Device Catalogue NumberSS7
Device Lot NumberDP-21634
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/09/2024
Supplement Date Received by Manufacturer01/16/2025
Initial Report FDA Received Date10/07/2024
Supplement Report FDA Received Date01/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-