The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The reported event was inconclusive because no sample was available for evaluation and further investigation did not result in any additional findings.Root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be ¿kinked tubing".However, there was insufficient information to confirm this potential root cause.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: proper techniques for urinary catheter insertion ¿ perform hand hygiene immediately before and after insertion.¿ insert urinary catheters using aseptic technique and sterile equipment.¿ use the smallest catheter possible, consistent with good drainage.¿ document the indications for catheter insertion, date and time of catheter insertion, and individual who inserted catheter.Indications for use: pvc & red rubber: for urological use only.Silicone: intended for use for bladder management including urine drainage, collection and measurement.The devices are generally passed to the urinary bladder via the urethra.Directions for use: 1.Wash hands and don clean gloves.2.Explain procedure to patient and open peri-care kit.3.Use the provided packet of wipes to cleanse patient¿s periurethral area.4.Remove gloves and perform hand hygiene with provided alcohol hand sanitizer gel.5.Using proper aseptic technique open csr wrap.6.Don sterile gloves.7.Place under pad beneath patient, plastic/ ¿shiny¿ side down.Note: use caution to maintain aseptic technique.8.Position fenestrated drape on patient.9.Saturate 3 foam swab sticks in povidone iodine.10.Lubricate catheter.11.Prepare patient with 3 foam swab sticks saturated in povidone iodine.Use the nondominant hand for the genitalia and the dominant hand for the swabs.Note: use each swab stick for one swipe only.Female patient: with a downward stroke cleanse the right labia minora and discard the swab.Do the same for the left labia minora.With the last swabstick cleanse the middle area between the labia inora male patient: cleanse the penis in a circular motion starting at the urethral meatus and working outward.12.Proceed with catheterization in usual manner using the dominant hand.A.When catheter tip has entered bladder, urine will be visible in the drainage tube 13.Document procedure according to hospital protocol.Bard® ez-lok¿ sampling port (indicated by the blue stem in the port) accepts luer lock (fig.1a) or slip tip collection devices (fig.1b).Bd vacutainer® urine products can be used for collection, storage, and transport of urine specimens from a urinary catheter.1.Swab surface of bard® ez-lok¿ sampling port with antiseptic (fig.2) e.G., site-scrub¿ ipa device which is an effective disinfecting agent for luer hubs.2.Using aseptic technique, position the collection device in the center of the sampling port.Press the device firmly and twist gently to access the sampling port.(fig.3) 3.If using bd vacutainer® luer-lok¿ access device (fig.4), collect urine into the bd vacutainer® tubes per hospital protocol.If using syringe, slowly aspirate urine sample into syringe and transfer to collection cup or follow hospital protocol.4.After sample is obtained, discard collection device according to hospital protocol.5.Follow established hospital protocol for specimen labeling and transport to lab.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|