It was reported that two ultrasonic surgical device used during a therapeutic celioscopy (dissection around a polypropylene element), had the scissor broke (lost jaws) in the patient's belly during adhesiolysis.After the second breakage, a monopolar hook and endobag were used to remove the broken pieces.The first device was used for about 30 minutes and the second one for about 10 minutes.There were no reports of patient or user harm associated with this event.
|
|
This report is related to patient identifier (b)(6) (1st device).The software installed was version 2.00.Intelligent tissue monitoring (itm) was turned on, is usually kept on, and was not changed during the procedure.The evaluation of the event/device is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's (lm) investigation.The device was evaluated by olympus, and the probe was confirmed to be broken.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than 1 year since the subject device was manufactured.Based on the results of the investigation, the probe likely broke due to the following mechanism: 1.The output was activated in seal & cut mode while the grasping section was grasping thick tissue.Therefore, the probe and the tissue pad came into contact at the rear end of the grasping section, causing the tissue pad to wear out.2.The non-insulated area of the grasping section and the probe came into contact due to wear of the tissue pad.3.The output in seal & cut was activated while the non-insulated area of the grasping section was in contact with the probe.As a result, a contact mark developed.4.A force to activate the output in seal & cut mode or a force to grasp tissue was applied to the probe.Therefore, cracks developed at a contact mark.5.A force was applied to the probe causing it to break.The event can be detected/prevented by following the instructions for use (ifu) which states: ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after the tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ ¿during the treatment, do not activate output while applying the probe tip to the tissue with a strong force, grasping thick tissue, or twisting the handle.Also, do not insert the handle while the handle is twisted with respect to the tissue, do not grasp it, and do not activate the output.Otherwise, the probe tip and/or grasping section may be damaged, which may result in falling of the probe tip and/or tissue pad.¿ this supplemental report includes a correction to g2 to provide information that was inadvertently not included in the initial medwatch.An update has been made to h3.Also, information has been added to h4.Olympus will continue to monitor field performance for this device.
|