The reported product is not expected to be returned as the customer was unwilling to return the device.An extended investigation has been conducted, which involved a manufacturing review (including device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation addresses the customer¿s complaint in this case and did not identify any trend or potential root cause that would indicate that the product is not meeting specification.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/deficiency.If a product defect/deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott's post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/risk ratio.No further investigation is planned.In the event that product is received, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
A battery/no power issue was reported with the abbott diabetes care (adc) device.The customer was unable to monitor their blood glucose due to the device not powering on with button press or test strip insertion.As a result, the customer was unable to monitor their blood glucose and experienced loss of consciousness, requiring treatment of "pancake syrup in finger" provided by non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|